What Is a Pure Steam Generator?
A pure steam generator is a clean-utility unit that produces steam free of pyrogens and particles from purified water. When the steam condenses, the condensate must meet the pharmacopoeial quality requirements for water for injection (WFI). This separates pure steam from plant steam, which comes from the boiler house and may carry treatment chemicals.
Where Is Pure Steam Used?
- Sterilisation in place (SIP) of vessels, bioreactors, piping and filter housings
- Autoclaves and steam sterilisers
- Sterilisation of filling lines and freeze dryers
- Cleanroom humidification
- Pharmaceutical, biotech, vaccine, dairy and personal-care production
Pure steam is expected wherever steam touches the product or product-contact surfaces.
How Does It Work?
Feed water, usually purified water, is evaporated in a heat exchanger heated by plant steam. In falling-film designs the water runs down the inside of the tubes as a thin film, which gives fast and efficient evaporation. The steam then passes a separator that removes droplets and the endotoxins they carry. A double tube sheet exchanger separates the heating steam from the pure steam side, so a leak cannot let boiler steam into the pure steam.
How Is Pure Steam Quality Tested?
Steam used for sterilisation is checked against three parameters defined in EN 285:
| Parameter | Limit | Why it matters |
|---|---|---|
| Non-condensable gases | max. 3.5 % by volume | Air and other gases block heat transfer |
| Dryness value | min. 0.90 (0.95 for metal loads) | Wet steam causes wet loads and poor sterilisation |
| Superheat | max. 25 K | Superheated steam behaves like dry heat and sterilises less effectively |
The condensate is also monitored for chemical purity and endotoxins against WFI requirements. The three physical tests can be run at the point of use with a portable pure steam quality test kit.
Selection Criteria
- Capacity: total steam demand in kg/h of all consumers running at the same time (autoclaves, SIP lines).
- Feed water quality: purified water or WFI feed, and any pretreatment needed.
- Materials and finish: AISI 316L, electropolished product-contact surfaces, orbital welding and sanitary fittings.
- Exchanger design: double tube sheet against cross-contamination.
- Automation: PLC control, alarms and interlocks, data logging, and 21 CFR Part 11 compliance where required.
- Compliance and documentation: CE/PED, ASME, qualification documents (DQ, IQ, OQ) and FAT/SAT support.
- Energy use: plant steam and cooling water consumption, heat recovery options.
Frequently Asked Questions
Are pure steam and clean steam the same?
The terms are often used interchangeably, but in pharmaceutical use pure steam means steam whose condensate is of WFI quality. Clean steam is a broader term and does not always guarantee that quality.
Can a pure steam generator also produce WFI?
Some designs can produce limited amounts of WFI with an added condenser. For continuous, high-volume WFI demand a multi-effect distillation unit is the usual choice.
How often should the quality tests be run?
The frequency follows the site’s validation plan. In practice they are run during qualification and then periodically, commonly at least once a year and after any significant change.
Solutions from EMA Engineering
EMA Engineering represents Pharmalab pure steam generators and clean-utility systems in Türkiye and supports selection, installation, commissioning and qualification.
- Pure Steam Generator – Pharmalab
- Pure Steam Quality Test Kit – Pharmalab
- Water for Injection Generation Plant – Pharmalab
- Autoclaves and GMP Sterilisers
Contact us for a capacity and configuration proposal for your project.